medical diagnostic equipment /cheap medical equipment MSLGA03A
Delivery Time: | 1 Days |
---|
Packaging: | Standard Export Packing |
---|---|
Brand: | MSL |
Place of Origin: | Guangdong, China (Mainland) |
Basic Info
Model No.: MSLGA03A, MSLGA03A
Product Description
Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of application | adult, child |
Gas source | O2: 0.27~0.55MPa |
Flow meter | O2: 0.1~2L/min 2.5~10L/min |
Flow rate of rapid oxygen supply | 25~75L/min |
Low oxygen pressure alarm(Oxygen Alarm) | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure,temperature,and flow rate.The regulation range of evaporator concentration is 0~5 vol%.Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer. |
Respiratory circuit | working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa |
Respiratory bellows | bellows for adults, bellows for children tidal volume range:0~1500 ml |
Ventilator (Ventilator Definition and Ventilator Function) | |
Display mode | Digital screen display |
Ventilation mode | IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate) |
Ventilation parameters | Tidal volume 50~1500ml Rate 2~99 bpm SIMV rate 2~20 bpm I:E 2:1~1:8 Inspiratory trigger pressure -1.0~2.0 kPa Pressure range 1.0~6.0 kPa |
Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform |
Security Alarm System | |
Airway pressure alarm | upper limit setting range 1.0~6.0 kPa low limit setting range 0.4~2.0 kPa |
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min low limit setting range 1.0~10L/min |
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input in 15s |
Power alarm Power |
AC 220V±10% 50Hz±1Hz |
Packaging & Shipping Safty Export Packing Client & Certificate Pls kindly contact: Contact us if you need more details on Medical Diagnostic Equipment. We are ready to answer your questions on packaging, logistics, certification or any other aspects about Cheap Medical Equipment、Medical Equipment. If these products fail to match your need, please contact us and we would like to provide relevant information.