Multi-channel X-Ray Fluorescence Element Analyzer


Energy dispersive X-ray fluorescence (EDXRF) is a well-established and powerful tool for nondestructive elemental analysis of virtually any material. X-Ray Element Analyzer is widely used for environmental, industrial, pharmaceutical, forensic, and scientific research applications to measure the concentration of elemental constituents or contaminants. The fluorescing atoms can be excited by energetic electrons, ions, or photons. A particular Single Channel Analyzer EDXRF method, monochromatic microfocus X-ray fluorescence (MμEDXRF), has proven to be remarkably powerful in measurement of trace element concentrations and distributions in a large variety of important medical, environmental, and industrial applications. When used with state-of-the-art doubly curved crystal (DCC) X-ray optics, this technique enables high-sensitivity, compact, low-power, safe, reliable, and rugged analyzers for insitu, online measurements in industrial process, clinical, and field settings. This new optic-enabled MμEDXRF Multi-Channel Element Analyzer technique, called high definition X-ray Fluorescence Element Analyzer (HD XRF), is described in this paper.

Multi-channel Element Analyzer


Principle of Analysis
XRF is used to analyze the element content in the material by using the X-ray fluorescence analysis technique and using the method of energy dispersion.



Product Features

♢ Real time: on-line real-time detection, and fast analysis results are given.
♢ Representativeness: the measuring device is well circulated and mixed evenly. The results of the measurement accurately reflect the real situation of the pulp.
♢ Adaptability: Measurements are not affected by site factors such as concentration, particle size, gas bubbles, velocity, delamination, etc. Using the normal temperature semiconductor detector, it does not need liquid nitrogen refrigeration and can work at normal temperature. The X-ray tube is used, and the radioactive source is not used.
♢ A new type of nuclear electronic technology and nuclear energy spectrum analysis technology have been developed to perform comprehensive analysis of complex energy spectrum and eliminate mutual interference between different elements.
♢ The working status is stable, the analysis accuracy is high, the reliability is strong, the operation is simple, the maintenance is convenient, and the software interface is friendly and humanized.



Product Parameters

1) Instrument size:Take six flow channels as an example, the instrument dimensions ( length × width × height): 4180x1650x2660mm;the net weight of six-channel X-ray fluorescent elemental analyzer:2450Kg;
2) Technical parameter:
♢ One analyzer can detect up to 12 flow channels;;
♢ Analytic element species: the atomic number of elements are over 20;
♢ content range of analysis element: typical content is 100% ~ 0.001%;
♢ Accuracy: The relative error of higher content is 1-4%, the relative error of lower content is 3-6%, and the relative error of extremely low content is 10-20%.
♢ Stability: Stability ≤ 0.3%, long-term drift can be automatically corrected.
♢ Analysis time: each sample flow is 1 minutes, and the time can be adjusted.
♢ Analysis cycle: analysis cycle of six sample flows is usually 8 minutes.
3) Sites condition
♢ Flow requirements: The optimal flow range for each set of analysis tanks: 5 to 10 m3/h.
♢ Irrigation water requirements: living water, clean and no suspended particles; 0.3MPa
♢ Power requirements: three-phase AC 380V, normal operating power is less than 4KW, ground wire to ground resistance is less than 4Ω.
♢ Compressed air requirements: Air pressure requirements: 0.6MPa
♢ Temperature requirement: working stream temperature: 0°C~40°C; working environment temperature: -20°C~50°C; working environment humidity: 0%~90%






Contact us if you need more details on Multi-channel Element Analyzer. We are ready to answer your questions on packaging, logistics, certification or any other aspects about Fluorescence Element Analyzer、X-ray Element Analyzer. If these products fail to match your need, please contact us and we would like to provide relevant information.